You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 13 Next »

Rotameter Head Loss Experiment

Size parameter must be a number (optionally followed by 'px', 'pt' or 'em').

Unknown macro: {float}
Unknown macro: {center}
Figure 1: Experimental Setup

Overview

A rotameter can be included between the alum stock tank and the constant head tank in the dose controller to verify the flow rate of chemical. The head loss through the rotameter needs to be determined in order to perform an accurate energy balance on the dose controller to provide precise dosing.

Experimental Setup

The head loss through the rotameter was be determined using the setup that is diagrammed in Figure 1. A peristalic pump was used to circulate the fluid at a certain flow rate. Two attenuators were used to minimize the effect of the pulsing from the peristalic pump, and one pressure sensor was used to measure the pressure difference before and after the rotameter.

A ramp function in process controller was used to control the flow rate of the circulating fluid. The flow rate was gradually increased from 8 ml/min to 380 ml/min.

The head loss was be determined by performing an energy balance around the rotameter. The pressure reading from the standing water is subtracted from the measured pressure difference values to cancel out the height term. The resulting pressure difference corresponds directly to head loss of the rotameter.

Results

One experiment looked at the response of the rotameter to flow rates of 8 to 380 mL/min. As seen in Figures 2 and 3, the experimental data fit best to a 0.09 in orifice with a 2 cm offset in pressure. This 2 cm of extra head needed to fit the 0.09 in orifice model has been hypothesized to be caused by the energy needed to lift the ball float.

Unable to render embedded object: File (3.09.10 Full Flow.jpg) not found.

Unknown macro: {center}
Figure 2: Full Flow Experiment #1 (3/09/10): 8-380 mL/min

Unable to render embedded object: File (3.11.10 Full Flow.jpg) not found.

Unknown macro: {center}
Figure 3: Full Flow Experiment #2 (3/11/10): 8-380 mL/min


In Figures 2 and 3 there is also a seemingly linear response in the data at low flow rates. Another experiment was performed at 8 to 40 mL/min flow rates to determine if this was indeed linear.

Unable to render embedded object: File (3.11.10 Low Flow.jpg) not found.

Unknown macro: {center}
Figure 4: Low Flow Experiment #1 (3/11/10): 8-40 mL/min


Unable to render embedded object: File (3.12.10 Low Flow.jpg) not found.

Unknown macro: {center}
Figure 5: Low Flow Experiment #2 (3/12/10): 8-40 mL/min


From the data, it seems that there is a linear response at low flow rates until about 25-30 mL/min.

Conclusions/Future Work

  • No labels