Autotutorial for Peristaltic Pumps

How do they work?

Peristaltic pumps are positive displacement pumps utilizing flexible tubes full of liquid and a rotor. Positive displacement means that a set amount of flow rate will be delivered to the system regardless of the pressure on either side. As the rotor turns, it squeezes the flexible tubing, forcing a specific amount of fluid out of the pump.

Pump Speed

The rotor on the peristaltic pump sets the pump speed by how many revolutions per minute that the rotor will turn. We have two types of pumps we currently utilize, one can vary between 1-100 rpm and the other can operate at up to 600 rpm.

Flexible Tubing Sizes

For each tubing size, different pump speeds will correspond to different flow rates being output by the pump.

The peristaltic tubing sizes are rather arbitrary and are labeled by numbers: 13, 14, 16, 17, and 18 in increasing order of size. There is also 3-stop tubing available, which are color-coded based on their ID and must be used with 6-roller pump heads.
The table below gives you an idea of the flow rates that can be achieved by a peristaltic pump for different tubing sizes.

 Tubing Code1314161718

3-Stop Tubing

Orange-Yellow

3-Stop Tubing

Yellow-Blue

3-Stop Tubing

Purple-White

 RPM/ID (mm)0.81.636.380.511.522.79

flow
rates
in
(mL/s)  

10.001000.003500.013330.046670.063330.0003240.00250.006167
1000.100000.350001.333334.666676.333330.03240.250.6166667
6000.600002.100008.0000028.0000038.00000---
mL/rev0.060.210.82.83.80.0190.1490.401

A Mathcad sheet with equations to help select tubing sizes and chemical stock concentrations is available for your use! 
From these charts, you can easily calculate a pump speed given a tubing number and flow rate. At the appropriate tubing number, divide the given flow rate by the flow rate at 1 rpm.

What tubing size and pump speed would you use to acquire a flow rate of 500 mL/min? Keep in mind that you can use more than one pump head...

Working with Microbore Tubing

When working with pump tubing of small size (#13, #14, Orange-Yellow, and Yellow-Blue), it is necessary to use microbore tubing to hydraulically connect the pump to the experimental apparatus. The microbore tubing most commonly used in the lab is PTFE with OD of 1.6 mm (1/16") and ID of 0.8 mm (1/32"). This tubing can be purchased at McMaster-Carr or Cole-Parmer. To connect the microbore tubing to the peristaltic pump tubing requires a barbed tube fitting sized for tubing of 1/16" ID  (can be purchased at McMaster-Carr), shown below.

This barbed tube fitting inserts easily into the small diameter pump tubing as shown below.

Connecting the microbore tubing to barbed fittings poses more of a challenge, since the barbed fitting is designed to fit inside a tube of 1/16" ID, but must instead fit over a tube of 1/16" OD. This would not be possible if not for the ductility of the microbore tubing. In stretching the microbore tubing, its diameter is decreased, and can be stretched to fit within the barbed fitting. Stretching can be accomplished in a number of ways, but the recommended approach is to use the puller shown below.

After coiling the microbore tubing tightly around its spool, the puller can be used to grip the microbore tubing and stretch it as shown below.

Stretching can also be accomplished by wrapping the tubing around the handle of a screwdriver, for example, in the absence of the puller.

Once the tubing is sufficiently stretched, locate the section of tubing where the cross section is narrowest. It is here that the tubing should be cut so that it can be pulled through the narrow opening. When cutting, try to make the cut at an acute angle, as shown below, so that the tubing will fit even more easily into narrow opening.

If the tube has been stretched enough and the cut has been made neatly, it should be fairly easy to pull the tubing through the narrow opening of the barbed fitting. Once the tubing comes out the other side of the fitting, pull it through until it fits snugly in the opening of the fitting, as shown below.

Once the tubing is secure in the fitting, any excess can be removed and the other side of the barbed fitting can be inserted in the pump tubing as shown below.

 

The same process of stretching the microbore tubing to fit it snugly in smaller openings can be applied in other places. One common application is to bore a 1/16" hole in a 1/4" push-to-connect plug so that the plug can then be inserted into a push-to-connect through-wall or tee fitting. This is the most common way to make the connection between microbore tubing and an experimental apparatus.


Pump Calibration

ProCoDA has a pump calibration routine available on the configuration tab. The pump must be on and running (controlled by ProCoDA) in order to complete the calibration.

 

  • No labels